BLOOD GLUCOSE METER OVERVIEW
This section provides general information on the Microdot Xtra Blood
Glucose Monitoring System
The Microdot Xtra Blood Glucose Monitoring
System is a portable whole blood testing
system which performs tests with quick
and accurate results.
It consists of three main parts:
1. Microdot Xtra Blood Glucose Meter
2. Microdot Xtra Test Strips
3. Microdot High and Low Control
Solution
These products have been designed,
tested and proven to work together
as a system to produce safe and
accurate results.
Use only Microdot Xtra Test Strips
and Control Solutions with the
Microdot Xtra Glucose Meter.
Your system includes:
Microdot Xtra Meter
Anti-microbial Meter Cover
Quality Assurance Manual
Warranty Registration Card
3V Lithium Battery (installed)
Additional components include:
Microdot Xtra Test Strips
Microdot Control Solutions
2
Chapter 1
Cambridge Sensors USA, LLC
®
Xtra
TROUBLESHOOTING THE METER
The blood glucose level is higher than
525 mg/dL.
The blood glucose level is lower than
20 mg/dL.
Error message that indicates that there
is a problem with the meter, e.g.,
measurement error (time out, overflow,
offset) or temperature out of range.
Error message that indicates that there
may be a problem with the test strip, e.g,
the test strip may be damaged, moved,
or removed during testing, or inserted
improperly.
Error message could be caused by a used
or damaged test strip.
Error message indicates serial
communications error.
The battery sign appears on the display
with the unit of measurement. The power
of the battery is getting low. You can
complete about 50 more tests from the
time this symbol first appears.
This message indicates very high
blood sugar. You should recheck the
blood glucose level. If “Hi” again,
follow facility protocol.
This message indicates very low
blood sugar. You should recheck the
blood glucose level. If “Lo again,
follow facility protocol.
Review the instructions and try again
with a new test strip. If the problem
persists, contact Customer Service.
Check the test strip for damage and
retest as necessary. Repeat the test.
If the error message appears again,
contact Customer Service.
Repeat the test with a new strip.
If the error message appears again,
contact Customer Service.
If error persists, contact Customer
Service.
Test results will be accurate but
replace the battery as soon as
possible. Battery type is
CR2032 3 volt.
Chapter 9
25
Cambridge Sensors USA, LLC
®
Xtra